Reprocessing and Sterilization

Caring for your instruments

Validated guidance on cleaning, lubrication, packaging and steam sterilization for your ErgoDenta instruments, aligned with EN ISO 17664, EN ISO 15883 and EN ISO 17665.

This guide applies to all reusable ErgoDenta dental and surgical instruments. Single-use items are marked as such and must not be reprocessed.

DPC Dental
Proud UK partners of ErgoDenta
EN ISO 17664EN ISO 15883EN ISO 17665
Based on ErgoDenta document ED-RPR-001, Rev 00

Read before processing

  • Use only trained personnel and validated procedures and equipment.
  • Never exceed 138°C (280°F). Do not use flash (immediate-use) sterilization.
  • Do not use phenols, strong acids or alkalis, iodophors or peroxide-based detergents.
  • Inspect every instrument after each cycle and replace any that are damaged or corroded.
  • Reprocessing applies to reusable instruments only. Disposable items must not be reprocessed.
1

Pre-Cleaning at the Point of Use

  • Within two hours of use, wipe off gross contamination with a cloth or swab moistened with 70% isopropyl or ethyl alcohol.
  • Keep instruments moist so that contaminants do not dry on.
  • Remove coarse debris with soft brushes only. Never use metal brushes or steel wool.
  • Open all hinged instruments fully.
  • ErgoTip modular instruments are processed assembled; there is no need to take them apart.
  • Transport instruments in a covered, puncture-resistant container.

Mouth mirrors and handles

  • Clean mirrors and handles separately.
  • Wipe mirror surfaces from side to side, not in circles.
  • Do not brush mirror surfaces. A soft brush may be used on the rest of the instrument if needed.
Back to top
2

Automated Cleaning and Thermal Disinfection

This is the recommended method. Use a washer-disinfector conforming to EN ISO 15883 (for example, Miele G 7836 CD) with Neodisher MediZym at 1%, or an equivalent neutral enzymatic detergent (approximately pH 6.5 to 8.5).

1%Detergent concentration
50°C · 10 minCleaning
93°C · 5 min+Thermal disinfection
110°C · 25 minDrying

Loading

  • Make sure all surfaces are exposed to the cleaning solution and water jets.
  • Keep instruments from touching one another.
  • Load hinged instruments open and unlocked.
  • Follow the washer manufacturer's loading instructions.
  • If any residue remains after the cycle, repeat the cleaning.

Washer requirements

The washer-disinfector should have approved efficiency, be regularly calibrated and maintained, run a suitable programme with adequate rinsing, and use good-quality water (deionised or purified where possible).

Suitability has been demonstrated under these parameters in accordance with EN ISO 15883.
Back to top
3

Inspection

After cleaning and before sterilization, check each instrument for:

  • residual contamination
  • surface corrosion or damage
  • smooth hinge and ratchet function
  • sharp, undamaged cutting edges
  • secure, clean ErgoTip threads
  • intact mirror surfaces.

Remove damaged instruments from service rather than repairing them. Contact DPC Dental for replacements.

Back to top
4

Maintenance

  • Sharpen scalers, curettes and cutters according to the manufacturer's instructions.
  • Check ErgoTip threads and replace them if worn.
  • Clean and lubricate any stiff or sticky hinges.
  • Replace mirrors when their reflectivity drops.
  • Record reprocessing cycles where local regulations require it.
Back to top
5

Lubrication (Hinged Instruments Only)

Use a medical lubricant that is compatible with steam sterilization, such as Sterilit Instrumenten Ölspray (300 ml).

  1. Apply lubricant to the hinge, box joint and ratchet.
  2. Open and close the instrument several times.
  3. Wipe off any excess with a lint-free cloth.
Back to top
6

Packaging

Make sure instruments are completely dry. Use STERIKING paper/film pouches (Wipak Medical) or equivalent double peel pouches that:

  • withstand at least 141°C (286°F)
  • allow steam to penetrate
  • protect the instruments from damage
  • meet EN ISO 11607-1.

Maintain and inspect heat sealers regularly.

Packaging steps

  1. Choose a pouch of a suitable size.
  2. Place hinged instruments in the pouch open and unlocked.
  3. Package ErgoTip units assembled.
  4. Heat-seal according to the manufacturer's instructions and check the seal visually.
  5. Label with the instrument name, sterilization date, expiry date and lot number.
STERIKING pouches have been laboratory-tested for steam sterilization in accordance with EN ISO 11607-1.
Back to top
7

Steam Sterilization

Use an autoclave conforming to EN 13060 (small sterilizers) or EN 285 (large sterilizers). Equipment must be validated and maintained in line with EN ISO 17665 and the manufacturer's instructions.

B-classFractionated pre-vacuum
134°CSterilization temperature
4 min+Holding time
30 min+Drying time

Do not use

  • Flash (immediate-use) sterilization for wrapped items
  • Gravity displacement cycles
  • Radiation, formaldehyde, ethylene oxide or plasma sterilization
  • Dry heat, unless the instrument's instructions for use permit it

Sterilizer maintenance

  • Follow the manufacturer's inspection and maintenance schedule.
  • Clean the chamber regularly.
  • Use only low-contamination, deionised or purified water.
  • Use an automatic drying programme if available, and confirm that items are fully dry.
Suitability has been demonstrated at 134°C for 4 minutes with at least 30 minutes of drying, in accordance with EN ISO 17665.
Back to top
8

Post-Sterilization: Cooling and Inspection

  • Let items cool fully inside the sterilizer, and do not compress warm pouches.
  • Check pouches for seal failures, tears or moisture, and reprocess any compromised pack.
  • Confirm that process indicators have changed correctly.
  • Check hinge and ratchet movement before storage.
Back to top
9

Storage

  • Store instruments in sealed packaging in a clean, dry, dust-free place, away from damage, humidity and sunlight.
  • Keep sterile and non-sterile items strictly separate and clearly labelled.
  • Do not exceed the pouch manufacturer's storage time or conditions.
  • Reprocess any item whose seal has been broken or compromised.
Back to top
10

Material Resistance: Incompatible Substances

Do not use detergents or disinfectants that contain:

  • strong alkalis (pH above 9)
  • strong acids (pH below 4)
  • phenols
  • interhalogens or iodophors
  • strong oxidisers or peroxides
  • organic solvents.

Never clean with metal brushes or steel wool. Chlorides and minerals in tap water can cause corrosion, so use demineralised or purified water where possible.

Back to top
11

Disposal

The instruments contain no hazardous materials, but they must be fully cleaned and sterilized before disposal. Follow local and national regulations for medical waste or reusable medical devices.

Back to top
Need help with reprocessing?

Talk to the DPC Dental team

We can help with validated cycles, instructions for use and technical files for CSSD audits, and advise on instrument management across your practice or group.